Hi Akash My Take on the difference is: For systems and equipment: Qualification is : Installation/Operational/Performance Qualification Protocols (Based Upon...
pauls@...
multivits2004
Offline Send Email
Dec 1, 2004
4:29 pm
12805
When equipment is relocated, especially Analytical Instrumentation, what is the consensus on revalidation of the instrument and/or any attached computer...
Joel Gates
biothrax
Offline Send Email
Dec 1, 2004
4:30 pm
12806
Dude, Assuming that it is the first signature you have to provide 2 components for signing.... Personnaly i prefer relying on the interpretation that signing...
emarniqu04
Offline Send Email
Dec 1, 2004
4:31 pm
12807
Hi Eric, I presume you're developing an application for use within the Pharmaceutical industry. If so, I think this is a good idea. You should make it so that...
hebburnbinman
Offline Send Email
Dec 1, 2004
4:31 pm
12808
Hi JG You would have raised a change request to get authorization to move said system, then under the due process a risk assessment of the impact of the move...
pauls@...
multivits2004
Offline Send Email
Dec 1, 2004
10:33 pm
12809
Hi Akash I've always found these definitions useful: Equipment is qualified Processes are validated. Don't remember where I first got them though :-( Wasn't...
John Dickson
john_dickson55@...
Send Email
Dec 2, 2004
12:36 pm
12810
Thanks Heb.... Actually we are already re-using the Windows session ID and password to restrict and control access to the application. Depending of the ...
emarniqu04
Offline Send Email
Dec 2, 2004
12:36 pm
12811
Depends on what you are moving, and where you are moving it to/from. I think that if you identify the equipment as being portable as part of the initial...
Dave Ward
david_ward99uk
Offline Send Email
Dec 2, 2004
12:37 pm
12812
Paul, I'm going to be controvercial here and disagree with your comments on Jargon Mismatch: The Software industry uses the term Software Validation, quite...
Dave Ward
david_ward99uk
Offline Send Email
Dec 2, 2004
12:37 pm
12813
Hi. The activities needed for a qualification an validation are the same. The differenc is the target. Qualification assure that the facility o equipment...
narcispv
Offline Send Email
Dec 3, 2004
9:38 pm
12814
A VSCR would be a good quick way to easily move equipment in a controlled manor. As Founds for calibration of let's say a spectrophotometer would be the first...
Richard Rence
r_rence_csv@...
Send Email
Dec 3, 2004
9:38 pm
12815
I don't find it hard to differentiate: = Qualification has a clear start and end. Can be managed, like a project. Applicable to any definable part of a...
Boeijen, F.P.M. (Frans)
boeijenf
Offline Send Email
Dec 3, 2004
9:38 pm
12816
Hi David I am taking part in ValBok. You have actually agreed with me in part, but what I keep saying is why does the software industry insist on using the...
pauls@...
multivits2004
Offline Send Email
Dec 3, 2004
9:38 pm
12817
Paul, You do a risk assessment. Is this just a section on the change Request? Is it done by one person? Richard what is a PQE? Andrew ... From: Richard...
Andrew Gibson
agibson@...
Send Email
Dec 6, 2004
1:57 pm
12818
Yes and while we are at it where did "Bespoke" come from why not "custom". Andrew ... From: pauls@... [mailto:pauls@...] Sent: Friday, 3...
Andrew Gibson
agibson@...
Send Email
Dec 6, 2004
1:58 pm
12819
Hi Paul I think testing is part of validation activities because validaiton is the documentation of testing activities as the evidence that the specific system...
Sakuracom
ikrimata
Offline Send Email
Dec 6, 2004
2:01 pm
12820
Paul, Agree, it's an unfortunate use of terminology which make it difficult to argue the case for validation in the Pharm industry. The user see's the vendors...
Dave Ward
david_ward99uk
Offline Send Email
Dec 6, 2004
2:01 pm
12821
Hi PaulS, David and others, I would also give my little cent into this discussion Validation vs. Qualification: 1.) I think that validation is a term...
Johannes Albrecht
ja200361
Offline Send Email
Dec 6, 2004
2:01 pm
12822
Andrew The change request is used to prompt an evaluation of what is to be done, is it a like for like change? (Yes ? OK then confirm it and carryon (maybe))....
pauls@...
multivits2004
Offline Send Email
Dec 6, 2004
7:36 pm
12823
Hi Johannes The FDA recently announced that it is now going to take part in the PICs convention (I think that is right). PICs has its own set of jargon based ...
pauls@...
multivits2004
Offline Send Email
Dec 6, 2004
7:36 pm
12824
Andrew I have had this one out a few times. Best to use a car as an example Custom: You could buy an off the shelf car and have certain options included,...
pauls@...
multivits2004
Offline Send Email
Dec 6, 2004
7:37 pm
12825
One of the poster's questions remains unanswered... Does anyone here have experience of an FDA inspection where either practice was the subject of a 483 or...
chemera2002
Offline Send Email
Dec 7, 2004
1:31 pm
12826
Hi, I have a question on data back up and recovery. The previous thread on this subject talks about validating back up in the VMP of each system. My question...
Andrew Gibson
agibson@...
Send Email
Dec 7, 2004
1:31 pm
12827
Very nicly put Paul, couldn't agree more. David ... Date: Tue, 7 Dec 2004 08:25:37 +1300 From: pauls@... Subject: RE: validation vs. qualification Hi...
Dave Ward
david_ward99uk
Offline Send Email
Dec 7, 2004
1:31 pm
12828
Hi, If there is a server hosting data for more than one "validated application" and there is a backup routine for that server, then that routine should ...
Mulero, Miguel
miguelmm89
Offline Send Email
Dec 7, 2004
6:20 pm
12829
Yes Eric, the application I saw also used the windows ID & password to electronically sign files. You have to prove the link between NT security and...
hebburnbinman
Offline Send Email
Dec 7, 2004
6:20 pm
12830
I think this is a good approach, however I didn't say that the software industry coined the term validation and therefore own it. My point was: software...
Johannes Albrecht
ja200361
Offline Send Email
Dec 7, 2004
6:21 pm
12831
Andrew, Generally for each validation system/application that you have, there should be documented evidence of a backup/recovery for it. If the...
eperaz@...
eperazzoli
Offline Send Email
Dec 7, 2004
6:25 pm
12832
Dear Group There has been a lot of discussion about software validation, as presented by the software industry. I would like to start a discussion as to the...
multivits2004
Offline Send Email
Dec 7, 2004
8:26 pm
12833
We revalidate instruments (HPLC, GC etc.) for every 6 months using the same test scripts (if there is any upgrade in the instrument the document will be...
part11guru
Offline Send Email
Dec 7, 2004
8:26 pm
Messages 12804 - 12833 of 19734   Oldest  |  < Older  |  Newer >  |  Newest
Advanced
Add to My Yahoo!      XML What's This?

Copyright © 2009 Yahoo! Inc. All rights reserved.
Privacy Policy - Terms of Service - Guidelines - Help