Dear Vamsi, Please see my answers below. I hope they resolve the
problem. Dylan
Dylan Williams
National Institute of Standards and Technology
dylan@...
www.boulder.nist.gov/dylan
[+1] (303) 497-3138
From:
mtt-11-on-wafer@yahoogroups.com [mailto:mtt-11-on-wafer@yahoogroups.com] On
Behalf Of vamsi_palukuru
Sent: Friday, October 17, 2008 4:59 AM
To: mtt-11-on-wafer@yahoogroups.com
Subject: [mtt-11-on-wafer] Regarding the Findlen--Command to input Low
frequency
1.
What is the command to input the low frequency in
Findlen.?
2.
You should see a softkey (F1-F12)
that does this.
2.Can these line standards be used to calibrate VNA using Wincal ?
As Cascade has reproduced the MultiLine TRL algorithm in WinCal,
I think that this should be possible.
3. What should be the length of OPEN standard. can it be arbitrary ?
Yes. The TRL algorithm only needs to know the length of the
reflect standard to within +/- 90 degrees.
Thanks,