1689
Yes,I want to tilt the old-style J-resolved 2D spectra. I solve this problem by pipe2txt.tcl to converse the data to .txt file, and then use octave to shift...
likenmr
Offline Send Email
Mar 3, 2008
2:36 pm
1690
Dear NMRPipe user Is there any way to add/substract two 2D spectra in NMRPipe environment? Regards, Veniamin...
nmn_crc
Offline Send Email
Mar 3, 2008
2:55 pm
1691
... Check addNMR, a utility distributed with NMRPipe. It can handle both of these functions along with many more --- Best regards, Kevin...
nmrkev
Offline Send Email
Mar 3, 2008
2:59 pm
1692
There is a macro shear.M in the $NMRPipe/nmrtxt directory. It is quite straightforward and you should be able to modify it to your needs. Cheers, Pedro...
Pedro Aguiar
pedaguiar13
Offline Send Email
Mar 3, 2008
4:11 pm
1693
I am a new user of NMRpipe. When I processed a 15N-noesyhsqc 3D experiment it looks folded. I could not find the apparent reason for this. Can any one help me,...
Hussan, Syed (NIH/NID...
hussans@...
Send Email
Mar 4, 2008
10:25 pm
1694
If a dimension is reversed, use "FT -neg". If a dimension has the first and second halves rotated, use "FT -alt". In some cases, especially for 13N noe data,...
delaglio@...
bigfdelaglio
Offline Send Email
Mar 5, 2008
5:28 pm
1695
I have recently installed NMRPipe on my Intel iMac running OS X 10.5.2 and have so far observed only a couple of problems. 1. The purple border is grey. 2....
joneilnmrpipe
Offline Send Email
Mar 6, 2008
10:50 pm
1696
Joe, For point 2 check out this previous post... http://tech.groups.yahoo.com/group/nmrpipe/message/1338 Specifically, you will need to execute the two...
Michael Parker Latham
mpl9700
Offline Send Email
Mar 6, 2008
11:46 pm
1697
Thanks Mike and Shaheen, that worked. Joe ... seem to remember they solved some of these types of problems....
joneilnmrpipe
Offline Send Email
Mar 6, 2008
11:52 pm
1698
I'd like to change my synthetic peaklist using the updateTab.tcl script. I'm not certain of the format for the -list/-data file that I'm assuming includes...
hallodb
Offline Send Email
Mar 7, 2008
2:02 am
1699
I am writing here to see if anyone else has noticed a discrepancy between O2P value calculated for NMRpipe processing of 3D spectra and the acquisition values...
max_gaetani
Offline Send Email
Mar 13, 2008
3:45 pm
1700
Hi, It seems that you have to adjust post-acquisition the 13C frequency by about 2.3 ppm in order to correspond to Wishart and Sykes. Why doe Bruker calculate...
Virgil Simplaceanu
vs2a@...
Send Email
Mar 14, 2008
12:29 am
1701
Hie Massimiliano! In many triple resonance experiments such as HN(CO)CA and HN(CO)CACB, there is a frequency list (fq2) to switch the 13C carrier from CO to CA...
ewen_lescop2001
Offline Send Email
Mar 14, 2008
11:27 am
1702
I'm running X11 1.1.3 - XFree86 4.4.0 (the Tiger version) with Mac OS X 10.5.2 (Leopard), and Victor's trick works perfectly. Thanks!...
m73682
Offline Send Email
Mar 17, 2008
3:24 pm
1703
Hi, In the Bruker standard spectrometer configuration, the 13C frequency reference is that of TMS, this is about 2.66 ppm downfield of DSS (see for example...
Maggie Johnson
majohnson_4
Offline Send Email
Mar 20, 2008
2:37 am
1704
Hi, I got a very old varian noesy pulse program. "noesy with variable d1 delay to match em filter". It's said to be able to reduce the total experimental time....
Jixun Dai
daijixun
Offline Send Email
Mar 20, 2008
5:05 pm
1705
I'm trying to swap the order of dimensions for processing with another application. I have a 4D noesy collected as: 1 2 3 4 H_Acc H_Don X_Don X_acc and I want...
geoffreymueller
Offline Send Email
Mar 27, 2008
12:59 pm
1706
HUZZAH!!! Greetings, colleagues. The various xyz2pipe/pipe2xyz flags -y -z -a do indeed re-order the data, not simply adjust the header. So also do the...
Frank Delaglio
bigfdelaglio
Offline Send Email
Mar 27, 2008
10:18 pm
1707
Dear nmrpipe group, When I process the experiment HMBC (pulse sequence "hmbcetgpl2nd"), acquired in Magnitud Mode. I get the expected spectrum but twice in...
Rosa Fayos
rfca_1
Offline Send Email
Apr 3, 2008
10:43 am
1708
Your -aq2D should be complex (States) in the conversion script. You could also try the FT -auto option in the indirect dimension. The rest should be okay. ...
Siddhartha Sarma
sidd@...
Send Email
Apr 3, 2008
11:26 am
1709
Thank you Siddhartha for your quick answer. I have tried setting States in -aq2D and FT -auto in the indirect dimension but I keep getting the spectrum...
Rosa Fayos
rfca_1
Offline Send Email
Apr 3, 2008
12:34 pm
1710
This can happen if you have the wrong number of D0 in your acquisition. The pulse programs contains 2 D0, is that what you have defined in the parameters? ...
Scott E. Norris
cberscot
Offline Send Email
Apr 3, 2008
2:00 pm
1711
Here's the usual way to convert and process Magnitude-mode spectra. -yN and -yT will generally be the same -yMODE Real -aq2D Magnitude During processing, use...
Frank Delaglio
bigfdelaglio
Offline Send Email
Apr 3, 2008
4:21 pm
1712
I have collected HNCACB and CBCACONH spectrum on varian 600 machine. the number of scan is set 32 for Nitrogen dimension and stoped before 31 scan is finished...
luzhenwei1977
Offline Send Email
Apr 3, 2008
8:33 pm
1713
Perhaps this section from the NMRPipe Reference Page will help ... ... If a given dimension of a spectrum is reversed, then that dimension should be processed...
Frank Delaglio
bigfdelaglio
Offline Send Email
Apr 3, 2008
9:04 pm
1714
Thank you. I tried to process with -neg/alt on carbon dimension and it didn't work. Any other suggestion? ... should be ... dimension....
luzhenwei1977
Offline Send Email
Apr 7, 2008
7:43 pm
1715
Is it possible that you may have different sweep widths between the two spectra? Cheers MIke ... -- Mike Osborne Iric-Institut de Recherche en Immunovirologie...
Mike Osborne
ozzy672003
Offline Send Email
Apr 7, 2008
9:18 pm
1716
Good afternoon, Give the example HNCACB scripts in our NMRPipe processing library a shot. That and other examples in the library should make the gcbcaconnh...
Ryan McKay
ryan_nanuc
Offline Send Email
Apr 7, 2008
9:49 pm
1717
Hi, I have acquired Noesy - C-hsqc spectrum on 900MHz Bruker spectrometer. The spectrum was acquired with the following pulse sequence: noesyhsqcetgp3d from...
michal.respondek
michal.respo...
Offline Send Email
Apr 8, 2008
1:32 pm
1718
I wonder whether the two experiments were done with the same array order of phases. The array parameter could be either 'phase2, phase' or 'phase, phase2',...
hsin
hzwang31
Offline Send Email
Apr 8, 2008
7:59 pm